- Hign-concerned Chemical: None
- NC or NO: Normally Open
- Structure: control
- Media: Air
- Material: Plastic
- Temperature of Media: Normal Temperature
- Pressure: Medium Pressure
- Model Number: Fa0520F
- Brand Name: ZhiRongHuaGuan
- Power Source: DC
- Origin: Mainland China
- Certification: CE

When power,N1 and N2 open into each other, N3 and N1 is disconnected, N3 and N2 is disconnected.
When no power,N1 and N3 open into each other, N2 and N1 is disconnected, N2 and N3 is disconnected.
|
Rated Voltage |
DC4.5V |
DC6V |
DC12V |
DC24V |
|
Voltage Range |
DC4.05V-4.95V |
DC4.8V-7.2V |
DC10.8V-13.2V |
DC21.6V-26.4V |
|
Rated Current |
450±15%mA |
375±15%mA |
185±15%mA |
96±15%mA |
|
Deflated Speed |
<3S(Discouraged Time Form 300mmHg to 15mmHg at 500CC tank) |
|||
|
Pressure Range |
0-350mmHg |
|||
|
Air Tightness |
<3mmHg/min(reducing pressure from 300mmHg at 500CC tank) |
|||
|
Insulation Resistance |
>50MΩ Min(DC 500V) |
|||
|
Insulation Class |
A |
|||
|
Life |
above 100,000Times |
|||
|
Working Fluid |
Air |
|||
|
Operating Temperature Range |
0°C-55°C |
|||
|
Power |
<2W |
|||
PRODUCT SPECIFICATION
1. Application:This specification applies to solenoid valves Fa0520F DC4.5V、6.0V、12.0V、24.0V.
2. Test condition:
2-1 The standard of environment test is the ambient temperature of 20℃ and relative humidity of 65%RH。If no disputes omlur,the test can be conducted under the below circumstance:the ambient temperature of 5-45℃and relative humidity of 30~80% RH.
2-2 position:Horizontal
2-3 Load:air 300mmHg
3. Pattern: Two position three-way
|
3-1 Rated voltage |
DC4.5V |
DC 6.0V |
DC 12.0V |
DC 24.0V |
|
3-2 Operating voltage |
DC4.05V-4.95V |
DC4.8V-7.2V |
DC10.8V-13.2V |
DC21.6V-26.4V |
|
3-3 Current |
450±15%mA |
375±15%mA |
185±15%mA |
96±15%mA |
3-4 Insulation class: A
3-5 Leakage: Pressurizing from 0 till 300mmHg in a 500CC tank. then holding for 30seconds.After than, measuring the dropped pressure value, which shall be less than 3mmHg/min
3-6 Exhaust speed:With 500CC volume, it takes less than 5seconds to exhaust the pressure from 300mmHg to 15mmHg
3-7 Dielectric Strength:It has to test between coil and outer mental parts at applying AC600V for 1second.(current≤0.5mA)
3-8 The lead tension: Hanging a 0.5 Kg static load on the terminal of the valve will not cause a damage on the valve.
4.Reliability test:
4-1 life:Turning the power on for 1 second and off for 2 seconds as one cycle; doing this action for more than 30,000 cycle. After the action, examining the product by test 3-5,3-6 and finding out performing well.
4-2 Low-temperature environment test:Under a temperature of 0℃ for 5 hours, then examining the valve by test 3-6,3-7and finding out it still perform well.
4-3 High-temperature environment test:Under a temperature of 50℃ for 5 hours, then examining the valve by test 3-5,3-6 and finding out it still perform well.
5.Others:
5-1 Fall test:Basic performance and the struture shall not be defected after following test; With a standard packaging conditions, then dropping it at the front, side, and edges part from the height of 50 cm to the concrete floor.
5-2 Unusual test: There is no burning and smoking occurrence while the power is on with Rated voltage for 8 hours continuously.